37702_ch.fm 12/10/07 1:57 pm
UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Device description
The Medtronic Model 37702 PrimeADVANCED Neurostimulator is part of a neurostimulation
system for pain therapy.
Package contents
Neurostimulator
■
Torque wrench
■
Product literature
■
Warranty card (USA only)
■
Registration form
■
Patient identification card
■
Patient identification card and registration
A patient identification card is packaged with this device. Advise the patient to carry the
identification card at all times.
c
The patient identification card packaged with the device is temporary; a permanent card
will be mailed to the patient when Medtronic receives the registration form.
The implant registration form registers the device warranties and creates a record of the device
in Medtronic's implant data system.
Device specifications
The neurostimulator is a multi-programmable device that delivers stimulation through 1 or more
leads. The stimulation settings are stored in programs. A program is a specific combination of
pulse width, rate, and amplitude settings acting on a specific electrode combination (up to 16
electrodes per program). Up to four programs can be combined into a myStim group. When
using more than one program, the pulses are delivered sequentially—first a pulse from one
program, then a pulse from the next program.
Pulse width, amplitude, and electrode polarity for each program within the group can have
different values. Rate, rate limits, ramping, and cycling for each program within the group have
the same values.
MA16388A004
Rev A
Medtronic Confidential
NeuroImplant_R01
37702 2007-12
English 5