Table 1. Operating values for the Model 37702 PrimeADVANCED neurostimulator
Programmable parameter
Cycling
Scheduled therapy
Group adjust
a
All values are approximate.
b
No more than 32 programs may be defined within the 26 myStim groups.
c
Only 100 mV resolution possible from 10.0 V to 10.5 V amplitude.
d
Rate limited to 85 Hz when three programs/ group are active and to 65 Hz when four
programs/group are active.
Table 2. Physical characteristics of the Model 37702 PrimeADVANCED
Description
Connector type
Height
Length
Thickness
case
connector
Weight
Volume
Power source
Storage temperature
Serial number model designator
Radiopaque Identification (ID) code
a
All measurements are approximate.
b
Hybrid combined silver vanadium oxide.
c
The serial number is the model designator followed by a number. The clinician programmer
displays the entire serial number beginning with the model designator.
MA16388A004
Rev A
Operating range and resolution
OFF, ON: 0.1 sec to 30 min (resolution: 0.1 sec from
0.1 sec to 1 sec, 1 sec from 1 sec to 1 min, 1 min
from 1 min to 30 min
OFF, ON: 1 to 8 events per 24 h period
OFF; ON
a
neurostimulator
Value
Octapolar, in-line 2.8 mm (0.110 in) spacing
65.0 mm (2.6 in)
49.0 mm (1.9 in)
15.0 mm (0.6 in)
15.0 mm (0.6 in)
67.0 g (2.4 oz)
3
39.0 cm
6.3 Amp hours, 3.2 V HCSVO
-18° to +52°C (0° to +126°F)
c
NKL
NKD
37702 2007-12
Medtronic Confidential
NeuroImplant_R01
a
b
cell
English 7