Section F2: Manual Door Lock Override - Nanosonics trophon2 Manual Del Usuario

Ocultar thumbs Ver también para trophon2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 150

SECTION F2: Manual Door Lock Override

Use ONLY when the probe is locked in the chamber and must be urgently retrieved.
Chamber surfaces may be hot and disinfectant present.
Gloves must be worn to avoid contact with disinfectant.
Do NOT manually open the door during a HLD cycle. Disinfectant mist will be present
and contact must be avoided.
1. Ensure the trophon2 is powered off.
2. Open the waste drawer, remove any liquid and obtain the key (Figure 15).
3. Locate the Manual Door Lock Opening Mechanism Cover on the right side of the trophon2 (Figure
16). Align the 4 key notches with the indents on the cover and turn the key ANTI CLOCKWISE to
unscrew.
4. Once the Manual Door Lock Opening Mechanism Cover is removed, align the 4 notches with the
grooves of the Manual Door Lock Opening Mechanism inside. Push and turn CLOCKWISE to 90
degrees to unlock the chamber door (Figure 17).
The probe is NOT DISINFECTED and CANNOT be used until it has completed a
successful HLD cycle or high level disinfected by an alternative method.
L03031 1.0 03/2018
Figure 15
Figure 16
21

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Trophon 2

Tabla de contenido