Environmental Conditions:
trophon2 Operating
temperature range: 17
27
Start Up from sleep
Menu
A1.2: Warnings
Hot Temperatures
Do NOT touch surfaces in the internal chamber. They can be hot and cause burns.
To avoid probe damage, ensure the probe is correctly positioned in the chamber. (See section
C3.3 for correct positioning of probe.)
Malfunctions
Do NOT attempt to open the chamber door during a cycle, power failure or system malfunction.
All repairs must be carried out by trained personnel.
Transporting the trophon2
Follow your facilities manual handling procedures for guidance on lifting heavy
objects.
The trophon2 weighs approximately:
Unpackaged 22kg (48.5lb).
Packaged 25kg (55lb).
If your trophon2 has been used, purge the trophon2 before transportation to remove the
disinfectant. (See section E2).
Electrical Device
Use the power cable supplied with the trophon2, connect to an earthed power outlet with the
correct voltage and frequency as specified on the product and in Appendix 1. Incorrect voltage
can damage the product.
The trophon2 must not be connected to the same circuit as a critical patient or life support device.
Spilled fluid can result in electrical shock. Avoid spilling fluids on or around the trophon2. Do not
immerse any parts of the trophon2 in liquid.
Do not attempt to access the internal mechanics. This may result in electric shock.