3.2.2.2
Evaluation as Phosphate PO
RES (resolution)
S0 (standard 0)
A0 (absorbance 0)
S1 (standard 1)
A1 (absorbance 1)
3.3 Programming procedure
Note: Each programming procedure has only to be followed once prior to the first test
evaluation.
Important Note: The following programming procedures are valid for the evaluation
form Phosphate-Phosphorus PO
the data mentioned in table
Note: After programming the calibration data, it could be helpful to mark the channel(s) 1
and/or 2 accordingly with a label.
3.3.1
LCK 348 Phosphate-Phosphorus PO
1.
Press POWER to switch on the instrument.
2.
Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3.
Select measuring range 1 or 2 for LCK 348 (0.5-5 mg/l PO
GREEN
Note: Press
4.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
5.
Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is
displayed.
6.
Press GREEN key. "-.---" is displayed.
7.
Press BLUE key until "--.--" is displayed. Press GREEN key. "S0" is displayed.
8.
Press GREEN key until "A0" is displayed.
9.
Press GREEN key until "S1" is displayed.
10. Press GREEN key. Enter "05.00" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "A1" is
displayed.
11. Press GREEN key. Enter "0.600" by using BLUE key to change value and
GREEN key to move to the next space. Press GREEN key until "Add" is
displayed.
Table 2 Evaluation as Phosphate PO
LCK 348
--.--
01.50
0.104
15.00
0.600
-P. To evaluate as Phosphate PO
4
3.2.2.2 Evaluation as Phosphate PO4 on page
key to toggle between measuring ranges.
4
LCK 349
--.--
00.15
0.090
04.50
0.728
4
4
LCK 350
--.--
06.00
0.104
60.00
0.753
please choose
4
-P
-P).
4
23.
23