Mode
Function
F-1
Setting Deviation Value Increase
F-2
Setting Alarm Value
F-3
Setting Measurement
Unit
F-4
Backlit ON/OFF
F-5
Laser setting
Save
Save and automatic shut off
Body Temperature Measurements
• Press the trigger to switch the meter on with one beeper. The device performs
a brief self test: all elements of the display are shown for approx. 1 second.
Following a successful self test with backlit (Red-Orange-Green),
• Press the "1 B/O" button for "Body temp" for accurate measurements in the
range of 32.0 to 42.9°C (89.6 to 109.2°F ).
• Hold the meter by its handle and point it toward the surface to be measured.
Measurement distance should be 5 to 15cm (2 to 6" ).
• Press the trigger to take a temperature reading. The temperature reading
appears in the display.
If the F-4 MODE of backlit & "2" button voice ON & F-3 MODE set alarm
temperature 38.0°C:
Temperature (°C)
T ≤ 32.0
32.1 ≤ T ≤ 37.5
37.6 ≤ T ≤ 37.9
38.0 ≤ T ≤ 42.9
T ≤ 43.0
Notice: Red Backlit is followed alarm temperature limit (Fever alarm) .
• Release the trigger and the reading will hold for approximately 20 seconds
after which the meter will automatically shut off
Note: Before shutting off, you can take further measurement.
• If the temperature goes above 109.2°F or 42.9°C, Hi will appear in the display.
If the temperature goes below 89.6°F or 32°C, Lo will appear in the display.
ENGLISH
"+"
0,1°C
Increase
0,1°C
°F
On
On
Temperature (°F)
T≤ 89.6
89.8 ≤ T ≤ 99.5
99.7 ≤ T ≤ 100.2
100.4 ≤ T ≤ 109.2
T ≤ 109.4
"-"
Default
Decrease
0°C
0,1°C
Decrease
38.0°C
0,1°C
°C
°C
Off
On
Off
Off
Fever Teller
Green
Green
Orange
Red
Green
Remark
Object mode
is not available
Efective Range:
-3°C~3°C
Only when the laser
head installed
Beep
Three beeps
One beep
One beep
Ten Rapid beeps
Three beeps
10