Ir Measurement Notes - Gima YI-400 Guía De Uso

Termómetro infrarrojo sin contacto
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Disposal: Follow the valid legal stipulations in respect of the disposal
of the device at the end of its lifecycle.

IR Measurement Notes

• Before measuring, be sure to clean surfaces that are covered with frost, oil,
grime, etc.
• If an object's surface is highly reflective, apply masking tape or flat black paint
to the surface before measuring. Allow time for the paint or tape to adjust to
the temperature of the surface it is covering.
• Measurements through transparent surfaces such as glass may not be accu-
rate.
• Steam, dust, smoke, etc. can obscure measurements.
• The meter automatically compensates for deviations in ambient temperature.
However, it can take up to 30 minutes for the meter to adjust to extremely
wide changes.
Body Temp
Object Temp
Body Temp range
Surface Temp range
Emissivity
Field of View
IR Spectral response
Display
Response time
Over range indication
Operating Temperature
Operating Humidity
Storage Temperature
Power Supply
Automatic Power Off
Weight
ENGLISH
Range
35.5 to 42.0°C (89.6 to 108.5°F)
In the remaining measurement range
0 to 100.0°C (32.0 to 212.0°F)
32.0 to 42.9°C (89.6 to 109.2°F)
0 to 100.0°C (32. 0 to 212.0°F)
0.95 fixed
D/S = Approx. 5:1 ratio
(D = distance; S = spot or target)
8 to 14 um (wavelength)
Backlit LCD display with function indicators
500ms
"Hi" or "Lo"
10°C to 40°C (50°F to 104°F)
10% to 90% RH operating
<80% RH storage
-25 to 55°C (-13 to 131°F)
(2) AA" Batteries
20 seconds (approx.)
181g
Accuracy
±0.2°C / ±0.4°F
±0.3°C /± 0.5°F
±0.8°C / ±1.5°F
12

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

25591

Tabla de contenido