Each cleaning and disinfection cycle includes a pre-cleaning step with
one wipe and a disinfection step with a second wipe.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact customer service at
1-888-307-8188 for assistance.
This device has been validated to withstand 5,000 cleaning and
disinfection cycles using the recommended disinfecting wipe/towelette.
The tested number of cycles is estimated by 1,000 tests per year on each
meter and around 2 cleaning and disinfection cycles per meter per day
for 5 years, the expect life of device.
Stop using the meter if you see any signs of deterioration such as the LCD
display is cracked or cloudy, the buttons no longer function, or the outer
casing is cracked. Please contact customer service at 1-888-307-8188
for a replacement meter if any of the signs of deterioration are noticed.
The meter should be replaced after the validated number of disinfection
cycles or the warranty period, which ever comes first.
NOTE:
• Do NOT clean or disinfect the meter while performing a test.
• Please follow the instructions on the package label of Micro-Kill+™ disinfecting
wipes for safe use of the wipes.
• If the meter is being operated by a second person, the meter and lancing
device should be decontaminated prior to handling by an assistant.
• Do NOT allow cleaning or disinfecting solution to get in the test slot, battery
compartment, or strip-ejection button.
• If you do get moisture in the test strip slot, wipe it away with the corner of a
tissue.
• Always dry the meter thoroughly before using it.
• Do not spray the meter directly with cleaning solutions especially those
containing water (i.e. soapy water), as this could cause the solution to enter
the case and damage the electronic components or circuitry.
32