Technical Specifications - ResMed Swift FX Nano Manual Del Usuario

Mascarilla nasal
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Technical Specifications

Pressure-flow curve
Dead space
information
Therapy
pressure
Resistance
Environmental
conditions
Gross
dimensions
8
The mask contains passive venting to protect
against rebreathing. As a result of
manufacturing variations, the vent flow rate
may vary.
Physical dead space is the empty volume of the mask to the
end of the swivel. Using the wide cushions it is 125 mL.
4 to 30 cm H
O
2
Drop in pressure measured (nominal)
at 50 L/min: 0.5 cm H
at 100 L/min: 2.0 cm H
Operating temperature: +41°F to +104°F (+5°C to +40°C)
Operating humidity: 15% to 95% relative humidity
non-condensing
Storage and transport temperature: -4°F to +140°F (-20°C to
+60°C)
Storage and transport humidity: up to 95% relative humidity
non-condensing
Small:
16.9" (H) x 3.94" (W) x 2.95" (D)
(430 mm (H) x 100 mm (W) x 75 mm (D))
Standard:
16.9" (H) x 4.09" (W) x 2.95" (D)
430 mm (H) x 104 mm (W) x 75 mm (D)
Wide:
16.9" (H) x 4.49" (W) x 2.95" (D)
430 mm (H) x 114 mm (W) x 75 mm (D)
Mask fully assembled with short tube assembly (no
headgear)
O
2
O
2
Pressure
Flow
(L/min)
(cm H
O)
2
4
20
8
29
12
37
16
43
20
49
22.5
52
30
61

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Swift nano for her

Tabla de contenido