Vaisala RFL100 Guía Rápida página 7

Ocultar thumbs Ver también para RFL100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 41
1. Select a suitable mounting location. A good location is easily accessible, protected from
water and condensation, and remains within the operating temperature range of the
RFL100 Data Logger:
• +2 ... +60 °C (+35.6 ... +140 °F) with alkaline batteries
• -20 ... +60 °C (-4 ... +140 °F) with lithium batteries
If you need to measure a wider temperature range, use a HMP110 or TMP115
probe and connect it using a cable. This way you can leave the data logger in
an environment that is suitable to its specification.
2. Attach the mounting bracket using one of the mounting methods shown in
(page
6). Orient the bracket vertically so that the probe or probe cable points down after
installation. Do not attach the RFL100 without the mounting bracket.
CAUTION!
a location where it would pose a hazard if dropped, ensure the mounting
bracket is securely fixed with screws or zip ties.
3. Slide the logger into the mounting bracket with the probe or probe cable pointing
downward.
4. Peel off the protective film from the display and remove the yellow plug from the probe.
5. If the probe is attached with a cable, mount the probe in the desired measurement
location and secure the cable.
6. Recommended: Apply location labels to the mounting bracket and the RFL100 Data
Logger according to your installation plan and company policy.
If you are mounting the data logger higher than 2 m (6 ft) or in
Figure 1
7

Publicidad

Tabla de contenido
loading

Productos relacionados para Vaisala RFL100

Tabla de contenido