INSTRUCTION MANUAL CODE 80202 REV A
Screen 7.1 Test procedure.
Screen 7.2
Instruction
«Sample» Button
Suction tube
On pressing «Sample» button,
the sample is suctioned.
Fig. 7.3. Sample entry for its
analysis.
J
.P. SELECTA s.a.u.
7 Performing tests
Select the test to be performed in the menu of "test overview", and select "OK" to
go to the menu of "test procedure" in the menu of "Test enter".
7.1 Test procedure menu
The "test procedure" menu is displayed as follows:
1. Factor window:
2. Result window:
3. Control:
4. Calibration:
5. Print:
6. Change No.:
7. Time:
8. Temperature:
9. Lamp:
10. Wavelength:
11. Instruction window:
12. Return:
13. Reaction window:
7.2 Running a test
To do a test, follow the instructions shown on the bottom right corner "instruction
window".
1. Instruction window: Aspirating distilled water
Put a tube with distilled water under the sample tube and press sample key (figure 7.3).
The measure result will be displayed on the "result window" and on the "reaction
window".
2. Instruction window: Aspirating reagent blank or aspirating sample blank
window
Hold the reagent blank or sample blank under the sample tube and press the Enter
key or the sample button. The programmed solution will be aspirated and measured.
3. Instruction window: Aspirating Sample
Hold the sample under the sample tube and press the sample button. The programmed
solution will be aspirated and measured.
4. Calibration: It will be described in Chapter 8
5. Control: It will be described in Chapter 9
7.3 Changing test
When changing over from one test to another, washing the flow cell is required to
prevent reagent carry over, causing wrong calibration and measurement result.
To wash the flow cell:
Place a container of distilled water under the sample tube. Press the Rinse key.
•
The aspiration/flow cell system will be washed continuously now, until the Rinse key
is pressed again. Let the flushing continue for about 30 seconds.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Feb/2012
(It can be modified without notice)
Factor
Test result
Perform the control test.
Perform the calibration test.
Print the test result.
Change the sequence number tested.
Display the time.
Display the temperature of the flow cell.
Show the status of the lamp: On/Off.
Show the wavelength selected.
Display the instruction information.
Return back to above menu.
A curve will be displayed to represent the reaction
Tel 34 937 700 877 Fax 34 937 702 362
Page: 31