Gima TOBI CLINIC Manual Del Usuario página 16

Ocultar thumbs Ver también para TOBI CLINIC:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

CLEANING THE MAIN UNIT:
To clean the device external parts always use a cotton cloth dampened with detergent. Don't use abra-
sive or solvent detergents.
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE
EQUIPMENT DO NOT GET IN TOUCH WITH LIQUIDS.
NEVER CLEAN THE EQUIPMENT WITH WATER.
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and
glasses) to avoid getting in contact with contaminating substances (after each utilization cycle of the
machine).
ACCESSORIES SUPPLIED
N°2 COMPLETE ASPIRATION JAR 2000ml
CONICAL FITTING
TUBES SET 8 mm x 14 mm
ANTIBACTERIAL AND HYDROFOBIC FILTER
FOOTSWITCH CONTROL (for versions equipped with footswitch control)
EUROPEAN POWER SUPPLY CORD
Replacing the antibacterial filter:
The filter is made of hydrophobic material that stops the passage of liquids into the same filter.
If you suspect the filter may have been contaminated and/or got wet or discoloured, always remove and
replace the filter.
If the equipment is to be used on patients with unknown pathological conditions or should you evaluate
the possibility of indirect contamination, remove and replace the filter after each utilization. The filter
is not designed for decontamination, disassembly and/or sterilization. If you suspect the filter may have
been contaminated and/or got wet or discoloured, always remove and replace the filter. If the equipment
is to be used on patients whose pathologies are known and not implying any indirect contamination risks,
we recommend to remove and replace the filter at the end of each work shift or else every month, even if
the equipment has not been used. 4000ml complete tank versions are available on request.
Aspiration jar: the mechanical resistance of the component is guaranteed up to 30 cycles of cleaning
and sterilization. Beyond this limit, the physical-chemical characteristics of the plastic material may show
signs of decay. Therefore, we recommend that you to change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment
of the said tube.
Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for
reuse. The component must be replaced if there are visible signs of decay of the material constituting
the said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked
to the employment of the said component. Therefore, after each cleaning cycle, it is up to the final user
to verify whether the fitting is suitable for reuse.
The component must be replaced if there are visible signs of decay in the material constituting the said
component.
Service life of the device: more than 10000-12000 hours of operation (or 3 years) in accordance with the
standard conditions of testing and operation. Shelf life: maximum 5 years from the date of manufacture.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine
should comply with ISO 10993-1 standards on material biocompatibility. The medical device is provided
without a specific suction probe. If this device must be used with a specific suction probe, the end user is
responsible for making sure it complies with the EN 10079-1 regulation.
CLEANING ACCESSORIES AND INTERNAL PARTS
Before using the device, the manufacturer advises you to clean and/or sterilize the accessories.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
• Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with con-
taminating substances;
• Disconnect the tank from the device and remove the said container from the support of the device.
• Separate all the parts of the cover (overflow device, washer).
eNgLIsh
16

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

28214282152821628231

Tabla de contenido