Recommended separation distance between portable and mobile radio-communication devices and the monitor
The surgical aspirator is intended to operate in an electro-magnetic environment where RF irradiated interferences
are under control. The client or operator of the TOBI CLINIC device can help prevent electro-magnetic interference
by keeping a minimum distance between the portable and mobile RF communication devices (transmitters) and
the TOBI CLINIC device, as recommended below, in relation to the radio-communication maximum output power.
Maximum nominal
output power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in
metres (m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum
nominal output power of the transmitter in Watt (W) depending on the transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied.
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by
the absorption and by the reflection from buildings, objects and people.
SYMBOLS
Caution: read instructions (war-
nings) carefully
Keep in a cool, dry place
Manufacturer
Product code
Medical Device complies with
Directive 93/42/EEC
WEEE disposal
Serial number
Fuse
ON / OFF
Separation distance from the frequency transmitter (m)
150KHz a 80MHz
3,5
√ P
d=
V
1
0,12
0,38
1,2
3,8
12
eNgLIsh
80MHz a 800MHz
12
√ P
d=
E
1
0,12
0,38
1,2
3,8
12
Follow instructions for use
Keep away from sunlight
Date of manufacture
Lot number
Type B applied part
Class II applied
Temperature limit
~
Alternating current
Atmospheric pressure limit
800MHz a 2,5GHz
23
√ P
d=
E
1
0,23
0,73
2,3
7,3
23