INSTRUCTION MANUAL CODE 80243 REV A
Figure 13
Figure 14
J
.P. SELECTA s.a.
January/11
Choose Unit of Standard Samples
Move the cursor onto "Unit" by pressing [»] and [«], followed by [ENTER], 8 units
are under your selection (Fig 13).
Move the cursor onto the unit you need, then press [ENTER]. You can also press the
number ahead of the unit directly by means of the numeric keypad.
If you don't want to select any item, press [RETURN] to go back to the last screen.
Set Quantity of Standard Samples
Move the cursor onto "Number" by pressing [»] and [«], followed by [ENTER].
(Fig. 14)
Input the number of your standard samples by means of the numeric keypad. After
pressing [ENTER], it returns to the setting screen.
Note:
Number range is from 1-9, other number is invalid.
Set concentration of standards
Before this step, you must blank first. Pull the blank cuvette into the light path in the
pre-testing interface, and press [ZERO].
Move the cursor to Concentration, and press [ENTER]. (Fig. 15)
Pull sample 1 in the light path, input the concentration of sample 1 by means of
the numeric keypad, then press [ENTER]. It goes into the next setting interface of