Battery Replacement; Guarantee - J.P. SELECTA UV2005 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80243 REV A
Fig. I
Fig. II
Fig. III
J
.P. SELECTA s.a.
Unscrew the two lamp chamber screws and remove the cover.
Attention
The cover could be very hot, remember to use gloves!
Connector
Unplug the 3 wire connector by pulling it up.
Loosen and remove the two white screws (Fig. I and II in the next image) from the
deuterium lamp.
Replace for a new pre-aligned lamp provided by SELECTA or an authorized technical
service.
Make sure that the lamp bridle wire coincides with the pintle position. Then, tighten
both screws.
Switch again the cable connector (make sure that the cable connection orientation
is the correct one.)
Switch the equipment on. After self-diagnosis, select 300 nm and press [ZERO].
Check that the light beam focuses the monochromator inlet slot.
As the lamp is a pre-aligned one, it will need just a minimum adjustment.
Unplug the equipment, cove the lamp chamber again and tighten both screws.
Install again the equipment cover. Pay attention not to pinch the cables during the
process.

8.2 Battery replacement

UV-2005 can save many data in the RAM memory, and so a battery is needed in the
circuit board.
You will have to replace the battery when it is exhausted.
Switch off and unplug the equipment
(VERY IMPORTANT: high voltage).
Unscrew the board's 13 screws and remove it.
Remove the old battery and place the new one (Fig. III)
Place again the board and the 13 screws.

9 Guarantee

This product is guaranteed for one year from the date of purchase and it covers
Battery
defective components and assembly.
Please keep the purchase ticket as it is the only valid document for the purchase date.
This guarantee does not cover damage caused by an incorrect use of the equipment.
This equipment must be properly validated before its use in the value sample analysis
for the user.
Any return without the authorization of J.P. SELECTA, s.a. is not accepted.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
January/11
(It can be modified without notice)
Page: 53
Tel 34 93 770 08 77 Fax 34 93 770 23 62

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4120020

Tabla de contenido