NESTLE SelectFlo COMPAT 9522041 Instrucciones De Operación página 8

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7
DIAGRAM 5
Completely assembled pump set.
Operating Instructions (cont.)
Plug power cord into a 120 VAC,
60 Hz. three wire grounded outlet.
Assemble Delivery Set
1. Open pump door by pushing/pulling
door latch.
2. Prepare and hang the delivery
set from an IV pole.
3. Insert delivery set drip chamber
into drip chamber guide (roller
clamp should be closed)
(Diagram 1).
4. Assemble silicone tubing around
the rotor (Diagram 2).
5. Insert pump adapter into bridge
and place tubing into tubing guide
(Diagram 3).
6. Close pump door.
Check Pump Settings
1. Press ON—alarm will sound
.
.
.
.
briefly and display
8
8
8
8. The pump
has memory, and will scroll through
the parameters and display the pre-
vious settings. It will return to flash-
ing the Infusion Rate (Diagram 4).
2. Check the Infusion Rate. To
p
q
change, press
(FASTER) or
(SLOWER) until desired rate is dis-
played.
3. Check Volume Delivered. Press VOL
DEL. until Volume Delivered is indi-
cated on the display panel. If
desired, this can be reset to zero by
pressing CLEAR.
Auto Filling
1. Open the roller clamp on the
SelectFlo delivery set. Remove the
luer adapter cap from the delivery
set, and place the cap in the luer
adapter cap holder available on the
pump.
2. Press the FILL key for automatic
priming of the delivery set. The
display will show the word FILL,
and the rotor will begin to turn to
indicate the unit is priming. Fill the
entire line. The prime cycle takes
approximately 2 minutes. The
pump will audibly alarm and flash
the infusion rate when filling is
complete. (Note: Automatic filling
may result in a small amount of for-
mula being expelled beyond the
luer adapter on the delivery set.)
Gravity Fill
1. Follow steps 1–3 listed under
Assemble Delivery Set.
2. Next, remove adapter cap
from the delivery set, and place the
cap in the adapter holder available
on the pump. Open the roller clamp
and fill the entire line with formula.
Close the roller clamp.
3. Proceed to steps 4–6 under
Assemble Delivery Set.
4. Complete steps 1–3 under Check
Pump Settings.
Begin Feeding
1. Connect the delivery set to the
patient's feeding tube.
2. Press START/STOP to start the
pump.
3. OFF will turn off the pump at any
time (Infusion Rate and Volume
Delivered are retained).

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Selectflo compat 199245

Tabla de contenido