Descargar Imprimir esta página

Promax SATHUNTER Manual Del Usuario página 26

Ocultar thumbs Ver también para SATHUNTER:

Publicidad

Idiomas disponibles

Idiomas disponibles

*CBR
Provides CBER measurement of the current test point.
COMMAND:
RESPONSE:
*VBR
Provides VBER measurement in DVB-S and LBER measurement in DVB-S2 of the
current test point.
COMMAND:
RESPONSE:
*TMP
Provides the internal temperature of the equipment
COMMAND:
RESPONSE:
*FRS
Provides / configures the frequency of the current test point. The configured value is
not stored in the E2P, therefore when the test point changes, the configuration
changes.
COMMAND:
COMMAND:
RESPONSE:
Page 8
USER'S MANUAL. REMOTE CONTROL COMMANDS SATHUNTER
*?CBR<CR>
*CBRyx.xxEyy<CR>
where y indicates whether the measurement floods above or
below the allowed margin.
"<" the measurement is smaller than the minimum measurable.
">" the measurement is larger than the maximum measurable.
" " the measurement is within the measurement margin.
where x.xx is the mantissa of the measured CBER.
where yy is the exponent of the measured CBER.
*?VBR<CR>
*VBRyx.xxEyy<CR>
where y indicates whether the measurement floods above or
below the allowed margin.
"<" the measurement is smaller than the minimum measurable.
">" the measurement is larger than the maximum measurable.
" " the measurement is within the measurement margin.
where x.xx is the mantissa of the measured VBER.
where yy is the exponent of the measured VBER.
*?TMP<CR>
*TMPxxxx<CR>
where xxxx is the internal temperature of the equipment in tenths
of Celsius degree.
*FRSxxxxxxx <CR>
where xxxxxxx is the new frequency in kHz..
*?FRS<CR>
*FRS xxxxxxx <CR>
where xxxxxxx is the frequency in kHz of the current test point.
06/2008

Publicidad

loading