Technical Specifications; Mechanical Specifications - Multimetrix MOH 700 Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Remark

Technical specifications

Resistance
measurement
Specifications
Electrical safety
Power supply
Safety fuses

Mechanical specifications

Dimensions
Mass
Accessories
Accessories supplied
with the instrument
milliohmmeter
Thermal effects
When measuring low resistances, you must take two parameters into account in
assessing the result:
- the measurement temperature,
- the test current and its duration of application (applying a high current to a
resistance for a long time causes its temperature to rise).
In effect, the temperature has a large influence on the measurement result and on
the value of the resistance measured. Moreover, the temperature coefficient of
most conductors is known.
For example, the temperature coefficient of copper is 0.4 % per ° C.
A copper conductor having a resistance of 10.00 mΩ at 20 ° C will have a
resistance of 10.40 mΩ at 30 ° C.
The measuring cables and the measuring instrument are also made of
conductors subject to the effects of temperature.
For the greatest possible accuracy, after the first measurement, make a second
with the measuring leads reversed.
The mean of the two results obtained will give an accurate, reliable result.
5 measurement ranges (selected using the rotary switch):
- 0...200.0 mΩ with a resolution of 100 µΩ - test current: 100 mA
- 0...2,000 mΩ with a resolution of 1 mΩ - test current: 100 mA
- 0...20.00 Ω with a resolution of 10 mΩ - test current: 10 mA
- 0...200.0 Ω with a resolution of 100 mΩ - test current: 10 mA
- 0...2,000 Ω with a resolution of 1 Ω - test current: 1 mA
Accuracy:
Resistance: ± 0.5 % of the reading ± 2 digits
Current: ± 0.1 %
Operating conditions:
-15 ° C to +55 ° C with the leads provided
Storage conditions:
-20 ° C to +65 ° C
Max. relative humidity allowed: 93 % RH @ 40 ° C
EN 61010-1, EN 50081-1 and EN 50082-1, IEC 68-2-29, IEC 1010, clauses
8.2, 8.3 & 8.4, IEC 68-2-1, -2-2, -2-3
Cat.IV, 20 V
8 x 1.5 V batteries (AA or equivalent)
- power supply circuit: 1.5 A, HBC, 5 x 20 mm, DIN
- current circuit: 1 A, HBC, 5 x 20 mm, DIN
- potential circuit: 0.5 A, HBC, 5 x 20 mm, DIN
250 x 190 x 110 mm
~1,500 g with batteries (exclusive of accessories)
MOH 700.............................................................................P06.2341.03
Supplied with:
- 1 carrying strap
- 4 separate 0.6m cables with crocodile clip ends (4-wire method),
- 2 x 0.8 m double cables with crocodile clip ends (2-wire method),
- 8 x 1.5 V batteries,
- 1 user's manual in five languages
English

Publicidad

Tabla de contenido
loading