Option
Description
SENSOR
Disable (off) or enable (on) warnings triggered by power
WARNING
failures.
OPTIONS
DEFAULT
Set the calibration to the factory default. A Default Cal
CALIBRATION
warning is entered in the alarm/warning log. Default Cal
is not recorded in the calibration history.
Start a calibration after the calibration is set the factory
default.
SAMPLE
Select the number of sample streams–1 (default) or 2.
STREAMS
Use the optional sample sequencing kit to add a
second sample stream.
CLEAR ALARM
Erase the alarms from the alarm/warning log and set
LOG
point alarm log.
SHOW VERSION
Show the firmware revision numbers for the main
NUMBERS
EPROM, internal neuron and external neuron.
SIO1 (or SIO2)
When an optional serial I/O module is installed, select
SETUP
the serial I/O module output format (computer or
printer) for measurement readings, alarms and
warnings. A maximum of two serial I/O modules can be
used (SIO1 or SIO2). If PRINTER is selected for the
output format, the options that follow are not available.
AUTO—Readings, alarms and warnings are sent out to
the computer as they occur.
QUERIED—Readings, alarms and warnings are sent
out to the computer when an RS232 query command is
sent from the computer.
TIME/DATE OUTPUT—When Yes is selected, the time
and date are sent out with readings, alarms and
warnings.
Appendix

Plumbing diagrams

Figure 9–Figure 12
show the plumbing connections for the analyzers.
Figure 9 Plumbing diagram – Alkalinity analyzer
1 Mixing chamber
7 Colorimeter drain
2 Rotary valve
8 Cleaner
3 Colorimeter
9 Reagent 2
4 Reaction coil
10 Reagent 1
5 Auxiliary valve
11 Standard 2
6 Pump (autoburette)
12 Holding coil
13 Standard 1
14 Mixing chamber
drain
15 Waste
16 Sample 2
17 Sample 1