Step 4
On pressing the ◀ button, the number
indicating the minute will start blinking.
Press the ▶ button to set the minute.
Step 5
Press the S button to finish and to enter the
'alarm2' mode.
Repeat steps 2 to 5 to set the remaining
time alarms (alarm 2 and 3).
Step 6
Press and hold the S button for 3 seconds to finish and turn the
meter off.
36 www.i-sens.com
Caring for Your System
• To minimize the risk of transmission of blood-borne pathogens,
the pre-cleaning and disinfection procedure should be
performed as recommended in the instructions below.
• Wash your hands thoroughly with soap and water after
handling the meter, lancing device, or test strips.
• If the meter is being operated by a second person who is
providing testing assistance to the user, the meter and lancing
device should be disinfected prior to use by the second person.
Pre-cleaning and Disinfection:
Pre-cleaning procedure is needed to clean dirt as well as
blood and other body fluids on the exterior of the meter and
lancing device before performing the disinfection procedure.
The disinfection procedure is needed to prevent transmission of
blood-borne pathogens.
• For the meter and lancing device, this pre-cleaning and
disinfection procedure should be performed once per week.
Note:
The life span of a KetoSens meter is 5 years. We
recommend disinfecting both the meter and lancing device at
least once per week. We have validated a total of 260 cleaning
and disinfecting cycles (260 Cleaning and 260 disinfection
cycles) to represent weekly cleaning and disinfecting over the
use life of your meter and lancing device.
1 pre-cleaning and 1 disinfection cycles per week * 52 weeks
per year * 5 years = 260 pre-cleaning and 260 disinfection
cycles
www.i-sens.com 37