⑦ Repeat the same procedure for the
lancing device (step ① to step ⑥).
* After the pre-cleaning and disinfection procedure, the
control solution should be tested to confirm that the
meter works properly before using the meter. Control
solution tests should be performed with two different
levels of control solutions (Level A and B). Verify that the
test results are within the range printed on the test strip
box. See pages 26–28 for how to do a control solution test.
Note:
If any of the following deterioration signs appear
after pre-cleaning or disinfecting, please stop using the
system and contact Customer Service at 1-800-429-5001
Mon–Sat, 9 am–9 pm EST. At all other times, contact your
healthcare professional.
• When the inscriptions on the exterior of the meter (or
lancing device) have been removed
• When the color of the meter (or lancing device) has
changed or faded
• When cracks or roughness develop on the meter (or lancing
device)
• When a part of the segment on the meter display does not
flash
• When control solution test results do not fall within the
stated range on the test strip box.
40 www.i-sens.com
Caution:
• Do not use other cleaners or disinfectants because other
chemicals have not been validated and may damage the
meter.
• Do not get fluids inside the meter through the test strip
port, data transmission port or battery compartment. Never
immerse the meter or hold it under running water because
this will damage the meter.
Caution: Storage and Handling
• Do not expose the meter to direct sunlight, heat, or excessive
humidity for an extended period of time.
• Do not let dirt, dust, blood, or water enter into the meter's test
strip port.
• Do not drop the meter or subject it to strong shock.
• Do not try to fix or alter the meter in any way.
• Keep the meter in a cool and airy place.
• Strong electromagnetic radiation may interfere with the
proper operation of this device. Keep the device away from
sources of strong electromagnetic radiation, especially when
measuring your blood β-Ketone.
• Store all meter components in the carrying case to prevent
loss.
Note:
You can get additional information or technical
assistance by calling our Customer Service Center: 1-800-429-
5001 (9 am–9 pm EST, Monday–Saturday). At all other times,
contact your healthcare professional.