TROUBLESHOOTING CHART F (OSD UNITS ONLY)
Compressor
ON
Other Symptoms
No OSD lights are illuminated.
No OSD lights are illuminated,
but red "Service Required" light is
illuminated accompanied by a
beeping audible alarm.
Both OSD lights are illuminated.
Yellow Low Oxygen light
is illuminated.
Yellow Low Oxygen light
is illuminated and an intermittent
audible alarm sounds every
five seconds.
*Refer to Specifications page for oxygen purity levels.
T R O U B L E S H O O T I N G
Power Light
ON
Possible Cause
Defective OSD.
Oxygen level is low*
Defective OSD
Oxygen level is low*
Oxygen level is low*
Possible Remedy
Check concentration with an oxygen analyzer.
If the concentration is within specification, replace
the PC board.
Check concentration with an oxygen analyzer.
If the concentration is within specification, replace
the PC board. If the concentration is low,
refer to low oxygen concentration
symptom in Troubleshooting Chart A.
Replace PC board.
Check concentration with an oxygen analyzer.
If the concentration is within specification, replace
the PC board. If the concentration is low, refer to
low oxygen concentration symptom in
Troubleshooting Chart A.
Check concentration with an oxygen analyzer.
If the concentration is within specification, replace
the PC board. If the concentration is low, refer to
low oxygen concentration symptom in
Troubleshooting Chart A.