Specifications - Sunrise Medical DeVilbiss Serie Manual De Servicio

Concentrador de oxígeno serie de 5 litros
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Catalog Number
Delivery Rate (Lower delivery rates
available for low flow applications)
Maximum Recommended Flow
Outlet Pressure
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 21°C only)
0-1500 M (0-4921 ft)
1500-3000 M (4921-9842 ft)
3000-4000 M (9842-13123 ft)
Operating Environment Range*
10°C to 35°C, humidity range
of 30% to 75%
Power Consumption
Weight
Sound Level (ISO 8359:1996 from front)
Dimensions
Pressure Relief Valve
Operating System
For units equipped with an OSD
the visible "low oxygen" indicator
will activate at the following level
Storage Conditions
Equipment Class and Type
Approval Body
and
Safety Standard
EMC Compliance To:
Typical Accumulator Pressure at
Maximum Recommended Flow
@ Sea Level
@ 750 M (2460 Ft)
@ 1500 M (4921 Ft)
*Note: The OSD performance at 10°C to 35°C, 95% R.H. through voltage range on the 515ADS verified at 670m.
Specifications subject to change without notice.
* 2
Note: EN ISO8359:1996 does not apply to 515AKZ.
* 3
Note: Wattage listed is for 50 Hz operation, wattage at 60 Hz will be higher.
S P E C I F I C A T I O N S
515ADS, 515ADZ
115V~, 60 Hz, 4.1 Amp
97-127V~, 60 Hz
385 Watts Average
515ADS only-320 Watts @ 2.5 LPM & below
(The audible alarm will alert at approximately 75%.
At less than 60%, the red "service required" light will activate.)
-40°C to 70°C, humidity range of 10% to 100%, including condensation
Class II Equipment Double Insulated;
CSA
CAN/CSA-C22.2 No. 601.1-M90
DEVILBISS 5-LITER SERIES
1 to 5 LPM
5 LPM
8.5 psig (58.6 kPa)
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at nominal voltage only:
No degradation in performance expected
based on tests of similar products.
Not recommended/Not tested
No degradation in performance across
the operating voltage range
(tested at sea level).
50 lbs. (22.7 Kilograms)
48.0 dbA overall average
27.75"H x 16"W x 14"D (70.5 x 40.6 x 35.6 cm)
40 psig±5psig (276 kPa±34.5 kPa)
Time Cycle / Pressure Swing
84% ± 2%
Type B Applied Part
IEC 60601-1-2
22-27 psig (150-190 kPa)
20-25 psig (138-172 kPa)
17-21 psig (120-145 kPa)
49
515AKS, 515AKZ
220V-230V~, 50 Hz, 1.7A; 230V~, 60 Hz, 2.4A
187-264V~, 50 Hz, 200V-264V~, 60 Hz
335 Watts Average
515AKS only-301 Watts @ 2.5 LPM & below
TUV
EN60601-1:1990+A1+A2+A13
* 2
EN ISO8359:1996
* 3
LT-1928