C O M P O N E N T T E S T I N G , R E P A I R , A N D R E P L A C E M E N T

PROPER REPAIR PROCEDURES

The DeVilbiss Oxygen Concentrator is designed for ease of
service.To aid service personnel, a Service Kit (part #444-501)
is available which contains the necessary gauges, tools, and
testing instruments to properly service the oxygen concentra-
tor. On parts that are sold separately, the part number is indi-
cated in parenthesis.
The following parts are included in the Service Kit:
1 Slotted bit
1 #1 Phillips bit
1 #2 Phillips kit
1 7/16" Socket l/4" Drive
1 Crescent wrench
1 8" Duckbill pliers
1 T-10 Bit
1 5/32" Allen bit
1 5/64" Allen bit
1 9/64" Allen bit
1 7/64" Allen bit
2 Presure/Vacuum gauge (part #PVO2D-601)
1 Tool box
2 Test Fittings (part #303DZ-637)
1 Torx screwdriver w/bits
1 AC/DC test light
1 1/4" Ratchet wrench
1 3mm Hexbit
1 T-15 Torx "L" wrench
1 10mm Socket l/4" Drive
1 1/4" Drive extension
1 Plastic storage case
1 Plastic error indicator tool
In addition to the Service Kit, an oxygen analyzer (part #O2ANA)
is needed to periodically check oxygen concentration levels.A volt-
meter will be needed for more accurate voltage testing.
NOTE: Be sure to read all of the steps involved before begin-
ning any of the procedures in this manual.
NOTE: After repairing or replacing a component run the unit for
20 minutes, check the oxygen concentration and test for leaks.
Test for leaks using a certified leak detection solution such as
SWAGELOK #MS-Snoop
ethylene glycol). Apply leak test solution to all fittings and hose
connections with the unit running. If an air leak is present, the
solution will bubble. All leaks should be repaired before putting
the concentrator back in service.
CAUTION: Do not apply leak test solution to any part
of the rotary valve or the main PC Board assembly.
WARNING:When servicing the DeVilbiss Oxygen
Concentrator, be absolutely certain that the correct
tools are used and that the parts are free of oil and
grease or any material not compatible with oxygen.
Teflon
®
tape is recommended and must be applied to
the male threads omitting the first thread to eliminate
LT-1928
®
or equivalent (must not contain
the possibility of tape particles entering the oxygen sys-
tem. LOX-8™ sealant may be used in place of Teflon
tape.
WARNING: Electric shock hazard. Do not remove cab-
inet.The cabinet should only be removed by a qualified
Sunrise Medical homecare provider.
WARNING: Disconnect the power cord from the wall
outlet before attempting repairs on the unit. Extra
care should be taken if it is necessary to operate the
unit with the cabinet removed.

CABINET REMOVAL

To remove front and back cabinets (Figures 2 & 3):
1. Ensure the unit is unplugged from the wall outlet.
2. Using a screwdriver, remove the six screws that hold the back
cabinet to the internal structure and the bib.
NOTE: All six screws are the same size.
3. Remove the back cabinet by sliding it toward the rear until clear.
4. Remove the front cabinet by pushing the top shoulders toward
the back of the unit, then outward away from behind the bib.
Tilt the top of the front cabinet forward until it can be pulled
out of the base of the unit.
The majority of all the servicing and repairs can be done with-
out removing the front bib. However, to gain access to the
components behind the bib, it may be loosened or removed.
To loosen the bib (Figure 4):
1. Remove the two screws (located directly above the hour
meter) that hold the bib to the unit's internal structure.This
will allow access to the components behind the bib.
To remove the bib completely (Figure 8):
1. Remove the two screws as above.
2. Disconnect the ribbon connector from the PC board.
3. Disconnect the lines from the power switch and circuit break-
er. Mark these wires accordingly.
4. Tilt the top of the bib forward to release it from the slot in
the body of the concentrator.
5. Remove the hose connected to the bottom of the flow meter.
To reassemble bib:
1. Reconnect the wires and hose.
2. Insert the bib tab into the slot above the rotary valve, and
push until it snaps into place.
3. Secure bib with two screws.
Teflon® is a registered trademark of DuPont.
LOX-8™ is a trademark of Fluoramics, Inc.
Snoop® is a registered trademark of SWAGELOK
14