Introduction

1.1 Outline

The Nikon non-mydriatic fundus camera RetinaStation is a fast and easy to use auto 3D tracking retinal
imaging system. RetinaStation is designed to provide images of the eye as an aid to clinicians in the
diagnosis of diabetic retinopathy, AMD, glaucoma and other retinal diseases.

1.2 Intended Use

RetinaStation provides non-mydriatic color retina and external images of the eye as an aid to clinicians in
the evaluation and diagnosis of eye disease. RetinaStation, which uses NIR LED as illumination when
aligning to the retina of a patient's eyes, only provides images and does not provide any diagnostic or
pathological analysis or conclusions in any way. Furthermore, RetinaStation is not intended to be used
as the sole diagnostic aid in disease identification, classification or management.

1.3 Proper Use of the Instrument

1. RetinaStation is a medical instrument; it must be operated by properly trained and qualified person(s)
only. Its operation should be supervised by a physician.
2. Please be sure to read the user manual and understand the safety precautions before operating this
instrument.
3. Always enter patient information first.
4. Prepare the patient contact surfaces (forehead rest and chin rest) according to the cleaning method
described in this manual.
5. Immediately turn off the power switch of this instrument and disconnect the power cable if unusual
problems arise.
6. Clean the ocular lens frequently to ensure good image quality.
7. Properly adjust the height of the motorized adjustable table to ensure the comfort of the patient
during examination.
8. Align the patient's eye position to the canthus indicator mark on the chin and forehead rest assembly.
9. Dim the room lights to allow natural dilation of the patient's pupil and provide comfortable viewing of
the fixation target without glare.
10. System functionality should be inspected once a year or whenever any repairs are made.
Page 2