[For data saved in Meter mode only]: To toggle display to
show absolute power in dBm or in both dBm & watt, press
F3 repeatedly.
To view saved time and date, press and hold Toggle
Centre.
To exit memory mode, press F1.
To retest or overwrite results
Note currently displayed memory location: At Home menu,
press button above "MEMORY". Note current memory
location e.g., "MEM 123".
Go to the memory location for retest/overwrite, select
desired location using + or -, then hold down button above
"MEMORY" and press F2 for three seconds or until it stops
beeping.
Press F1 to exit Memory mode.
To retest: Press F3 (Meter mode) or F4 (Autotest mode) to
return to data acquiring mode. Press button above [STORE]
now will save new results in the memory location selected
above.
Return memory location pointer for next storage to where it
was before the retest: Press button below [HOME] then that
above "MEMORY", select the desired location ("MEM 124"
in this example) using + or -, then hold down button above
"MEMORY" and press F2 for three seconds or until it stops
beeping.
Continue testing.
Clearing Memory
To clear all stored data, in Memory mode, hold down button
above "MEMORY", then press and hold F2 followed by F4
until it stops beeping.
Transfer Instrument Memory Data to USB Stick:
At Home menu, press button above "MEMORY" to enter
memory mode.
Insert USB stick into USB port ("MEM") on the right side of
the instrument and, press triangular button above "USB".
Wait till instrument stop beeping.
Pass/Fail Operation:
Loss test status for each wavelength will be shown with a
'
' (pass) or 'X'(fail) or blinking 'X' (marginal) on LCD. ORL
(2-way Autotest only) test status for each wavelength will be
shown with a '
' (pass) or 'X'(fail). A failed length test value
(2-way Autotest only) will flash.
Overall test result, "PASS" will be shown at the top right
corner of the if LCD:
-
pass/fail check function is disabled.
-
loss test statuses of all wavelengths are pass or
marginal and, both the optional length and ORL tests
pass.
-
An overall test result, "FAIL" will be shown at the top
right corner of LCD if, loss test status of any wavelength