Kingfisher KI23400 Serie Guía De Referencia Rápida página 6

Medidor bidireccional de pruebas de pérdida
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 40
is a fail or, any of the optional Length or optional ORL
test fails.
Pass/fail method selection:
For 2-way Autotest: If 'WC' is selected, the worst-case loss
value will be used to determine the test status. If 'AV' is
selected, the average loss value will be used to determine
the test status.
For 1-way Autotest: Worst case loss value will always be
used to determine the test status regardless of the selected
pass/fail method.
Pass/Fail Threshold Set Up:
To downloaded from KITS reporting software
Refer to instruction in the software manual.
To set up directly on the instrument
Turn on instrument with "-" button (for SM), or "+" (for MM)
key depressed.
"Imported Setup" will show if a setup has been imported from
KITS and will vanish if any changes are made locally.
Press F3 to toggle between loss / ORL setup screens. For
instruments without Length and/or ORL options, set the
corresponding threshold values to "0".
Press Toggle Centre repeatedly to select a value to adjust
threshold for PASS and/or MARGINAL.
Use - or + to adjust value. Set values (magnitude) of
MARGINAL must always be greater than PASS.
Use F4 to toggle WC (worst case) / AV (average) pass/fail
method.
Press F2 to save or F1 to cancel new settings.
Tone Detection Operation:
Connect 2 instruments with a fiber i.e., at the source-port on
one unit and the meter-port on the other.
On source unit, enter Source menu, press F3 to scroll
modulation options.
On meter unit, it detects the standard optical test tone (e.g.,
270 Hz, 1 KHz, 2 KHz), a buzzer sounds and the display
changes to show the actual measured modulation frequency
in Hz.
To end the operation (on source unit), press and hold F3,
then press Toggle Centre. Alternatively, press and hold F3
for 3 sec.
Slow Mode Operation:
To enter Slow Mode during turn on, hold down F3 & press
On/Off. The flashing symbol '
has been activated.
Alternatively, to turn slow mode on/off while in Meter mode,
press and hold key F3 then press F4.
To turn Slow Mode on while in Tone detection mode, press
F4. This will disable tone detection.
To turn Slow Mode off (i.e. to resume Tone detection), press
F3 to go to Meter menu, hold F3 then press F4.
' indicates that slow mode