Method
GB
Photometer COD vario
Select the appropriate vial for the desired range:
LR-Range:
0 –
MR-Range:
0 – 1500 mg/l, Order code 2420721
HR-Range:
0 – 15 000 mg/l, Order code 2420722
Preparing the Sample
Open a reaction vial with a white cap and add the specified volume.
(Ensure that appropriate personnel safety equipment is used.)
LR-/MR:
2 ml water sample
HR:
0.2 ml water sample
Prepare a blank (Note 1) by using deionised water (TOC-free) instead of the sample
(LR/MR: 2 ml, Hr: 0.2 ml).
Replace the cap tightly. Invert the vial gently several times to mix the contents (The vial
will become hot during mixing!) and digest the vials for 120 minutes in the reactor at
a temperature of 150 °C. Remove the vials from the reactor and allow them to cool down
to 60°C or less. Mix the contents by inverting each vial several times while still warm. Then
allow the vials to cool to ambient temperature before measuring.
On
Off
Lr
Mode
RANGE
Zero
Test
RANGE
0.0.0
12
150 mg/l, Order code 2420720
Measurement
Fix the adapter for 16 mm vials on the sample chamber.
Switch the unit on using the [ON/OFF] key.
The display shows the following:
Select the required test using the [MODE] key:
LR
MR
HR
AbS1
The display shows the following:
Place the blank in the adapter (Note 1 – 4) making sure that the
marks ı are aligned. Place the cover on the adapter.
Blanks are specially prepared for each individual test range.
Press the [ZERO/TEST] key.
The "Range" symbol flashes for approx. 8 seconds.
The display shows the following:
After zeroing remove the vial from the adapter.
Place the sample in the adapter (Note 2 – 4) making sure that the
marks ı are aligned. Place the cover on the adapter.
AbS2
LR
......... (Scroll)