Removing The iFill Oxygen Cylinder

WARNING
1.
Do not modify any connections on the personal oxygen station. Never lubricate the connections. Never use tools of any kind to connect/disconnect the oxygen cylinder
and the oxygen station. Otherwise, severe injury and/or damage may occur.
2.
Use extreme care when handling and filling an oxygen cylinder. Full oxygen cylin ders are under pressure and can become a projectile if dropped or mishandled.
NOTE – Refer to the G
D
eneral
1.
Remove the full oxygen cylinder by pushing down on the fill port collar located under the cylinder nipple connector
while using your other hand to steady the cylinder (Figure 4).
2.
Lift up on the cylinder to remove from the oxygen fill connector. The green Standby light will turn on after a short delay.
3.
Oxygen will escape from the fill connector for a short time after the cylinder has been removed. This may be
accompanied by a "pop" with a rush of air. This is normal.
4.
Perform one (1) of the following steps:
a.
If desired, fill another oxygen cylinder. Refer to the DeVilbiss iFill Personal Oxygen Station Operation Checklist.
b.
Place the oxygen fill connector cover onto the fill connector.
NOTE – The oxygen fill connector cover should always be used whenever cylinders are not being filled.
Setting The iFill Oxygen Cylinder To The Prescribed Setting
1.
Attach the nasal cannula to the cannula fitting of the oxygen regulator or oxygen conserving device. Open the oxygen
cylinder valve.
2.
Attach the cannula to your nose/face.
3.
Turn the rotary selector to the setting prescribed by your physician.
4.
Breathe normally.
WARNING
Changing the setting on the rotary selector will affect the dose of oxygen delivered, DO NOT readjust the setting unless directed by your physi cian.
5.
To turn the flow of oxygen off, turn the rotary selector counterclockwise to the "OFF" position. Close the oxygen cylinder valve.

TROUBLESHOOTING

The following troubleshooting chart will help you analyze and correct minor system malfunctions. If the suggested procedures do not help, call your DeVilbiss homecare
provider. Do not attempt any other maintenance.
WARNING
To avoid electric shock hazard, do not remove the cabinet. The cabinet should only be removed by a qualified DeVilbiss homecare technician.

DeVilbiss iFill Personal Oxygen Station

Problem
No indicator lights on.
Red light on.
Excessive fill times OR Green
filling light stays on.
Fuse Replacement (models with fused AC inlet connector)
1.
Disconnect AC power cord from AC inlet connector.
2.
With a small screwdriver, gently open the fuse drawer on the AC inlet connector.
3.
Remove the bad fuse from the exposed end of the fuse drawer, discard the bad
fuse.
4.
Remove the replacement fuse from the fuse drawer and slide it into the exposed
end of the fuse drawer where the bad fuse was removed.
5.
Insert the fuse drawer into the AC inlet connector until it is fully seated and flush
with the AC inlet connector housing.
SE-535I
& W
and the H
anGers
arninGs
anDlinG
Solution
1. Check that the personal oxygen station power cord is properly plugged into the wall outlet.
2. Ensure the fuse in the Mains Power Entry Module is not open.
3. If there are still no indicator lights, contact your DeVilbiss provider.
1. Ensure the iFill oxygen cylinder is connected correctly.
2. Press the start/stop button. If the Red light remains on, contact your DeVilbiss provider for service.
3. If the Red light turns on during the filling process, check that the cylinder is connected properly. Remove the cylinder and reinstall.
Ensure the cylinder rotary selector is in the OFF position. Press the start/stop button to turn off the Red light and then again to restart
the cylinder filling process. If the Red light turns on again during the filling process, contact your DeVilbiss provider for service.
1. Ensure the iFill oxygen cylinder rotary selector is set to OFF.
2. Check the fill times shown in Typical Oxygen Cylinder Fill Times chart.
3. Check that the cylinder is connected properly. Press the start/stop button to return to Standby. Remove the cylinder and reinstall.
Press the start/stop button to begin filling the cylinder.
4. If problem persists, contact your DeVilbiss provider for service.
W
in this manual.
arninGs
Figure 4
EN - 9