Troubleshooting; Specifications - Heidolph REAX 20 Instrucciones De Empleo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

E
Work on electric and electronic components is limited to qualified personnel.
No shaking
 check power cord
 check circuit breakers
 Thermal motor circuit breaker triggered by motor overload
Remedy:
Wait about 20 minutes, decrease load applied on shaker plate.
 Mechanical parts broken (humming motor noise) or electronic failure (no motor noise).
 Micro-fuse (K) blown, change fuse.
REAX 20
Power supply
Input power
Shaker frequency
Speed control
Speed display
Mixing movement
Drive
Capacity
Max. bottle diameter
Max. bottle height
Standard bottle
Dimensions (I x b x h)
Weight
Protection class
ambient temperature

TROUBLESHOOTING

SPECIFICATIONS

230/240 V 50/60 Hz / 115 V 50/60 Hz (optional)
280 W
1 – 16 1/min
electronically, continuously variable,
analog scale
overhead rotation
capacitor motor, motor overheating protection with
automatic reset
Reax 20/4
- 4 bottles
Reax 20/8
- 8 bottles
Reax 20/12 - 12 bottles
136 mm
270mm
a plastic wide necked bottle with a volume of 2000 ml.
Along with the proper accessories, the 500 ml and
1000m (sizes are also applicable)
REAX 20/4: 490 x 465 x 520 mm
REAX 20/8: 770 X 465 x 520mm
REAX 20/12: 1050 x 465 x 520 mm
Reax 20/4
23 kg
Reax 20/8
28 kg
Reax 20/12 33 kg
IP 21
0°C to 40°C at 80% rel. humidity
Approved for installation in gassing and conditioning
cabinets (make reference to temperature limits)
20

Publicidad

Tabla de contenido
loading

Tabla de contenido