3. Setting time to zero (0:00) and continuous mode: Accumulated time.
a. Press and hold the on/off button below the time display. After three (3) seconds,
the display will indicate the previous set time.
b. Simultaneously press both the up and down arrows, the display will indicate
zero (0:00). The unit time is now set to zero (0:00) minutes. Alternately, you can
use the up/down arrows to get to zero (0:00).
c. Press the on/off button below the time display. The display will indicate the
accumulated time. The up/down arrows will become inactive. To stop timer,
press the on/off button again. IMPORTANT: This will NOT interrupt the stirring
function. Press the on/off button below the speed display to interrupt the stirring
function.
d. To re-set, press and hold the on/off button below the time display. After three (3)
seconds, the display will indicate the previous set time, which was zero (0:00).
4. Setting timed mode: Programmed time.
a. Press the up/down arrows below the time display until you reach the desired
time.
b. Start this function by pressing the on/off button below the time display. The unit
will run for the selected time, the up/down arrows will become inactive while the
timer is running. The unit will stop stirring when the time display reaches zero
(0:00). Four (4) audible beeps will indicate the time down function is complete.
The time display will default back to the set time. To repeat for the same time,
simply depress the on/off button again.
c. To interrupt an automatic timing cycle before it is completed, press the on/off
button below the time display. The display will flash off and on to indicate the
time function is on "hold". IMPORTANT: This will NOT interrupt the stirring
function. Press the on/off button below the speed display to interrupt the stirring
function. Restart the timer by pressing the on/off button below the time display,
unit will continue counting down to zero (0:00). When the display reaches zero
(0:00), you will hear the four (4) audible beeps that indicate the time down
function is complete and stirring function will cease.
v
s
- sTHvDg
olume
Tirrer
5. Turning unit off:
a. To turn the unit off, press the standby button. The speed and time displays
will be blank, the standby indicator light will illuminate. The stirrer should be
kept in standby mode when not in use. To completely cut off power to the unit,
disconnect the power cord from the unit or un-plug from the wall outlet.
OPERATING TIPS
Built-in memory maintains the last used speed and time settings during a power
interruption.
The stirrer increases speed at a steady rate until the set-point is reached, if the stir
bar is too large or the liquid is too viscous, the stirrer may not reach its set-point. The
set-point speed needs to be reduced. Use stir bar supplied with the unit for optimal
performance.
The following errors should NOT be address by the user. If any of these errors appear,
contact your Ohaus representative for repairs.
T
- H
v
roublesHooTing
igH
olume
The following errors should NOT be address by the user. If any of these errors appear,
contact your Ohaus representative for repairs.
Error Code
Software Test
E03
motor system failure
s
- sTHvDg
Tirrer
Cause
- rotor locked
- motor failure
- motor sensor failure
- magnetic material placed near or on the
top plate surface
6