PRECAUTIONS

4
PRECAUTIONS
A rubber dam should be used with any dental procedure.
Always unplug the unit before changing fuses or adjusting the voltage
selector.
Applying too much pressure with the Flow hand piece may result in a
broken cannula.
Place the cannula gently into the canal. Too much pressure will stop
the motor from turning.
Take care when replacing a cartridge during a procedure, the cartridge
nut and the expended capsule may be hot to the touch.
The forward portion of the Flow hand piece becomes warm during use.
The heat shield (included with the system) may be used optionally to
reduce the hand piece surface temperature. If the heat shield is not
utilized, avoid contact with the forward portion of the Flow hand piece.
When replacing tips during a procedure, the plugger tips may be hot to
the touch.
The Calamus device may be used only with original Calamus tips and
cartridges!
This product must only be used in hospital environments, clinics or
dental offices by qualified dental personnel.
Chemclaving is not recommended for sterilization and maintenance of
the tips as this would cause corrosion.
Do not stay in the root canal with a heated tip for more than 4 seconds
for safety!
Place gutta-percha on the tip of the thermal response tip before
applying the tip to the patient's tooth! Do not place the tip directly on
the tooth!
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)