Basic Operation
Power Sensitivity Adjustment: This dial is normally set fully counter-
clockwise, which enables the UPS to provide maximum protection against
waveform distortions in its AC input. When such distortion occurs, the UPS
will normally switch to providing sine wave power from its battery reserves
for as long as the distortion is present. In areas with poor utility power or
where the UPS's input power comes from a backup generator, chronic
waveform distortion could cause the UPS to switch to battery too frequently,
draining its battery reserves. You may be able to reduce how often your UPS
switches to battery due to moderate waveform distortion by experimenting
with different settings for this dial. As the dial is turned clockwise, the UPS
becomes more tolerant of variations in its input power's AC waveform.
Note: The further the dial is adjusted clockwise, the greater the degree of waveform
distortion the UPS will allow to pass to connected equipment. When experimenting with
different settings for this dial, operate connected equipment in a safe test mode so
that the effect on the equipment of any waveform distortions in the UPS system's
output can be evaluated without disrupting critical operations.
Input Breakers: The SMART2500XLHG uses two input breakers to protect
your electrical circuit from overcurrent draw from the UPS load. One breaker
protects LINE, while the other NEUTRAL. If these breakers trip, remove some
of the load and reset them by pressing the breakers in.

Guidance and Manufacturer's Declaration

Product Information
AC Input Voltage
Input Amp(s)
No. of Phases
Class Type
Plug/Connector Type
Operation
AC Input Frequency
AC Input Protection (Qty./Type/Rating)
Max AC Leakage Current (uA)
Weight
Contraindication
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
This Medical-Grade UPS is intended for use in the electromagnetic environment specified below. The
customer or the user of this Medical-Grade UPS should assure that it is used in such an environment.
Emissions Test
RF Emissions CISPR 11
RF Emissions CISPR 11
16-07-007-933607.indb 10
120V
16A
Single
Class 1
IEC C20 inlet
Continuous
60Hz
Input breaker (2x 20A thermal)
Less than 100uA
73.3 lb. (33.2 kg)
None
Compliance Electromagnetic Environment - Guidance
Group 1
The Medical-Grade UPS uses RF energy only for its internal
function. Therefore, its RF emissions are very low and unlikely
to cause any interference in nearby electronic equipment.
Class A
The Medical-Grade UPS is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
10
8/31/2016 10:04:09 AM