INSTRUCTION MANUAL CODE 80221 REV 0
Fusible
Cable de red
Main switch
Used water
outlet
Fig 4.2
Filling the water
Door lock
tank
Fig 4.3
J
.P. SELECTA s.a.
June/07
4 Installation
Once the unit is unpacked, visually check, the packing list (See 1.1) and that there
are no parts damaged during transportation.
Save the packing for some days for later sending.
4.1 Suitable location
Place the unit over a fl at level and stable surface, suitable for the unit weight.
No longer that 1.5m from a power supply plug.
No tap water and waste will be needed.
To work comfortably, let a free space of 50cm on each side.
To allow ventilation let and not least than 20cm on the rear part.
Do not block or cover any ventilation holes.
Prepare to refi ll the water reservoir from the top of the unit.
4.2 Connection to power supply
Locate a power supply socket (Rated to 16A) near to the unit.
4.3 Filling water reservoir.
Use distilled or deionized water with this autoclave.
Use the supplied funnel to refi ll the tank.
Fill the tank with approx. 1.5 litre . See Fig 4.3
4.4 Connection to used water tank
Use the supplied hose to connect the used water tank to autoclave.
Connect the hose from the water outlet to the used water tank inlet. See Fig 4.2.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
(It can be modifi ed without notice)
For safety reason the socket must have ground terminal,
since the unit must work grounded.
Do not manipulate the power cord with wet hands.
Before to plug on the socket check that its voltage match
the voltage marked on the unit plate.
Plug the power cord on the autoclave rear part.
Used water tank
Tel 34 937 700 877 Fax 34 937 702 362
Page: 11