Control Panel - J.P. SELECTA AUTESTER P 1.5 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CODIGO 80221 REV 0
J
.P. SELECTA s.a.
Junio/07

5 Control panel

The control panel allows function selection and shows operation status.
Press main switch to power on the autoclave. See Fig 4.2
Display Pressure/
• Display pressure /min:
• Display temperature:
• Status indication:
Power:
Water:
Ready:
Running:
Load:
Warm-up:
PP1, PP2:
PP3:
SP:
EP:
END:
ERROR:
Error table.
Item Error Code
1
E1
2
E2
3
E3
4
E4
5
E5
6
E6
7
E7
8
E8
9
E9
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modifi caciones sin previo aviso)
Display tempe-
time
rature
Status
Pressure (bar) or remaining time.
Temperature (ºC) on the chamber
Autoclave status indication:
Unit powered On.
Water level on the reservoir.
Lights ON when water must be added. (Add water)
Ready to start. (Press START)
Cycle running
Ready to load material to sterilize
Warm up. (Up to temperature reach 100ºC)
Thermal vacuum. (Display indicates pressure)
Rise to sterilization temperature. (Display indicates pressure)
Sterilization. (Display indicates time to fi nish)
Draining water
Cycle end. Open can be opened
Error on the operation, See table
Failure
Temperature sensor error
Over pressure (>2.6bar).
Chamber over temperature.
Keep warm keep pressure.
Exhaust failed (20s< 0.3bar)
Door open on working time.
Steam generator Overtemperature
Pre-heating fault >20min
Filling steam gen. fails >30min
Selected cycle
Cycle selection
Stop/start
cycle
Action to take
Check sensor
Check pressure sensor
Check temperature sensor
Check valve
Check door look
Check temperature sensor
Check cylinder heaters
Check steam pipes
Tel 34 937 700 877 Fax 34 937 702 362