10
1.4 EN/IEC 60601-1-2:2007 sub-clause 5.2.2.2 Table 6
Recommended separation distances between portable and mobile RF communications equipment and the Coltolux®LED unit .
The Coltolux® LED unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled . The
customer or the user of the Coltolux® LED unit can help prevent electromagnetic interference by maintaining a minimum distance
between the portable and mobile RF communications equipment (transmitters) and the Coltolux® LED unit as recommended below,
according to the maximum output power of the communications equipment .
Rated maximum output power
of transmitter in watts (W)
0 . 0 1
0 . 1
1 . 0
10
10
For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer .
NOTE 1 At 80Mhz and 800MHz, the separation distance for the higher frequency range applies .
NOTE 2 These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorbtion from reflection from
structures, objects and people .
Separation distance according to frequency of transmitter in meters (m)
150kHz to 80MHz
d = 1 . 2 √ P
. 1 2
. 3 8
1 . 2
3 . 8
12
80MHz to 8000Mhz
d = 1 . 2 √ P
. 1 2
. 3 8
1 . 2
3 . 8
12
Owner's Guide
800MHz to 2 . 5 GHz
d = 1 . 2 √ P
. 2 3
. 7 3
2 . 3
7 . 3
23