Select-A-Flow* device is available in two flow rate ranges and distinguished by color on the
face of device:
Green (1-7 ml/hr)
• Flow rate dial - 1, 2, 3, 4, 5, 6, 7 ml/hr
White (2-14 ml/hr)
• Flow rate dial - 2, 4, 6, 8, 10, 12, 14 ml/hr
INSTRUCTIONS FOR USE
Use Aseptic Technique
FILLING THE EASYPUMP
PUMP (FIGURE 5)
Note: Follow hospital protocols and applicable
regulations for filling pump.
1. Close clamp.
2. Uncap the fill port.
3. Attach filled syringe to fill port. Invert pump
as shown.
4. Grasp syringe with both hands.
5. Push down on plunger continuously until
volume is dispensed. Do not handle pump
while filling, as the syringe tip may break.
Repeat as necessary. Syringe accuracy is
±4%.
Note: Filling Extension Sets are provided with
larger pumps (see product insert).
6. Remove filling device from fill port
7. Replace fill port cap. Label with appropriate pharmaceutical and patient information.
FILL VOLUME
CAUTION: Do not underfill pump. Underfilling the pump may significantly increase the
flow rate. Do not exceed maximum fill volume. (Table 1)
Table 1: Fill Volume (ml)
Labeled Fill Vol.
270 ml
400 ml
C-BLOC* RA™
®
Maximum Fill Vol.
335 ml
550 ml
Figure 5
Retained Vol.
≤11 ml
≤16 ml
6