Curve
Arrow
Ruler
Make Calibration to determine the corresponding relation between magnification and resolution, this
will establish the corresponding relationship between measurement unit and the sensor pixel size.
Calibration needs to be done with the help of a micrometer. For detailed steps of carrying out calibra-
tion please refer to chapter 8.4
Export the measurement information to CSV file(*.csv)
Delete All the Measurement Objects
Settings
Exit from Current Measurement Mode
Notes:
1. When user left-clicks Display/Hide button
case the Camera Control Panel will not pop up automatically even if moving mouse cursor to the left side of the video window.
Only when users left click the
doing other operations in the Camera Control Panel, the Auto Focus Control Panel or the Synthesis Camera Control Toolbar.
8.4
Camera calibration
The camera must be calibrated before any measurements can be taken.
To do this it is necessary to use the micrometer slide supplied with the camera.
Detailed procedure is the following:
1. Insert the objective with the lowest magnification available and put the stage micrometer on the stage.
2. Click on the "Calibrate"
3. The following dialog box appears:
on the Measurement Toolbar, the Measurement Toolbar will be fixed. In this
button on the Measurement Toolbar to exit from the measurement mode they will be able to
button in the Measurement toolbar.
Page 11