16. Microphotography

16.1 Use of "C"mount camera

1. Loosen the clamping screw ① on the trinocular port and re-
move the dust cap ②. (Fig. 65)
2. Screw the C-mount adapter ③ to the camera ④ and insert
the round dovetail of the C-mount into the empty hole of the
trinocular port, then tighten the clamping screw ①. (Fig. 66)
16.2 Use of Reflex camera
1. Insert the Reflex adapter ① into the relay tube to the micro-
scope ②.
2. Screw the "T2" ring ③ (not provided) to the reflex adapter.
3. Connect the Reflex camera ④ to the "T2" ring just installed
(Fig. 67).
4. Mount the other end of the relay tube ② into the empty hole
of the trinocular port, then tighten the clamping screw. (Fig.
65)
"T2" ring is not provided along with the microscope, but is
commercially available.
While shooting dark specimens, darken eyepieces and view-
finder with a dark cloth to minimize the diffused light.
To calculate the magnification of the camera: objective magni-
fication * camera magnification * lens magnification.
If using an SLR camera, mirror movement may cause the
camera to vibrate.
We suggest lifting the mirror, using long exposure times
and a remote cord.