13. Microphotography

13.1 Use of C-mount camera

1. Loosen the clamping screw
move the dust cap②. (Fig. 36)
2. Screw the C-mount adapter
the round dovetail of the C-mount into the empty hole of the
trinocular port, then tighten the clamping screw ①. (Fig. 37)
13.2 Use of Reflex cameras
1. Insert the Reflex adapter
scope ①.
2. Screw the "T2" ring
3. Connect the Reflex camera
(Fig. 38)
4. Mount the other end of the relay tube
of the trinocular port, then tighten the clamping screw. (Fig.
36)
"T2" ring is not provided along with the microscope, but is
commercially available.
While shooting dark specimens, darken eyepieces and view-
finder with a dark cloth to minimize the diffused light.
To calculate the magnification of the camera: objective mag-
nification * camera magnification * lens magnification.
If using an SLR camera, mirror movement may cause the
camera to vibrate. We suggest lifting the mirror, using
long exposure times and a remote cord.
on the trinocular port and re-
to the camera
into the relay tube to the micro-
(not provided) to the reflex adapter.
to the "T2" ring just installed.
into the empty hole
and insert