Smoke Evacuation System, 220 VAC
System Description
The CooperSurgical Smoke Evacuation System 6080A-50 three-stage air
filtration system is used to remove airborne particulate plume produced
during office and surgical procedures and has the following features –
• Low noise level.
• Triple filtration of air provides efficiency level for 0.014 microns rated
at 99.999%. This includes a pre-filter, a charcoal filter for odor
removal and a final safety filter placed after the charcoal filter.
• Adjustable high air flow for effective collection of plume.
• Conveniently attaches to the CooperSurgical LEEP 1000
WorkStation—wall mounted or free-standing.
Cautions
• U.S. Federal law restricts this device to sale by or on the order
of a physician.
• This device produces a strong vacuum force and care should be
exercised to ensure that the suction control and the position of the
inlet end of the suction tubing are adjusted properly to prevent injury
to the patient or inadvertent damage to surgical materials.
• The materials removed from the plume by this device are potentially
hazardous. Handle according to 29 CFR 1910.1030 and OSHA
3127.1992 (Occupational Exposure to Blood Borne Pathogens)
guidelines.
• To prevent a fire or explosion hazard, do not use the system in the
presence of flammable or potentially flammable materials.
• Do not allow fluid to be pulled into the system.
• To prevent premature failure of the ULPA filter cylinder, do not
operate this device without a disposable pre-filter in place.
Initial System Set-Up
You have received one carton which contains the Vacuum Unit. The
disposable accessories necessary to conduct procedures are to be
purchased separately from CooperSurgical.
English
1
®