Masimo® rainbow® Tester
Reusable
PURPOSE
The Masimo® rainbow® Tester is used to verify the general functionality of Masimo devices or devices containing Ma-
simo rainbow SET® Pulse CO-Oximetry. The Tester allows a connection to Masimo rainbow SET devices. Each device
manufacturer is responsible for determining whether its devices are compatible for use with the Tester.
CONTRAINDICATIONS
The Tester is not to be used as a calibration tool for pulse CO-Oximeters.
WARNINGS
• The Tester should be free of visible defects. Never use a damaged product, or one with an open enclosure shell or
exposed electrical contacts.
• Failure to properly connect the Tester to the device or patient cable will result in intermittent readings, inaccurate
results or no reading.
• Functional Testers cannot be used to assess accuracy.
INSTRUCTIONS
A. Attaching the Tester to the Device or Patient Cable
Orient the Tester such that the connector mates with the connector port on the device or patient cable and insert
firmly.
B. Taking a Test Measurement
1.
Turn the device on.
2.
A properly connected Tester will cause the device to display readings.
3.
The resulting representative data points should be as follows:
SpO
2
Pulse Rate = 61 bpm ± 1 bpm.
SpCO® = 7% ± 3%. *
SpMet® = 3.5% ± 3%. *
SpHb® = 14.0 g/dl ±1 g/dl. *
* If applicable
NOTE: SpO
and SpMet will alarm at factory default alarm values.
2
NOTE: If the displayed values differ from the specified values, contact Masimo Technical Support. Contact
information can be found at www.Masimo.com/globalservices
4.
Remove the tester from the device or cable.
CLEANING
Clean the Tester by wiping with 70% isopropyl alcohol and allowing it to dry thoroughly.
CAUTION
Do not use petroleum-based or acetone solutions, or other harsh solvents for cleaning. These substances affect the
devices materials and device failure can result.
WARRANTY
Masimo warrants to the initial buyer only that these products, when used in accordance with the directions provided
with the Products by Masimo, will be free of defects in materials and workmanship or a period of six (6) months from
the date of purchase. The manufacturer's sole obligation under this warranty is to replace any product that it deems to
be covered under warranty with a replacement Tester.
WARRANTY EXCLUSIONS
This warranty does not extend to any product that has been subject to misuse, neglect or accident; that has been
damaged by causes external to the product; or that has been used in violation of the operating instructions supplied
with the product. The warranty does not extend to any product that has been connected to an unlicensed device system,
modified accessories, or any unit that has been disassembled or reassembled by anyone but an authorized agent.
IN NO EVENT SHALL MASIMO BE LIABLE TO BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT, SPECIAL
OR CONSEQUENTIAL DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED OF THE
POSSIBILITY THEREOF. IN NO EVENT SHALL MASIMO'S LIABILITY ARISING FROM ANY PRODUCTS SOLD TO BUYER
(UNDER A CONTRACT, WARRANTY, TORT OR OTHER CLAIM) EXCEED THE AMOUNT PAID BY BUYER FOR THE LOT OF
PRODUCT(S) INVOLVED IN SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY DAMAGES ASSOCIATED WITH
A PRODUCT THAT HAS BEEN REPROCESSED, RECONDITIONED OR RECYCLED. THE LIMITATIONS IN THIS SECTION SHALL
NOT BE DEEMED TO PRECLUDE ANY LIABILITY THAT, UNDER APPLICABLE PRODUCTS LIABILITY LAW, CANNOT LEGALLY
BE PRECLUDED BY CONTRACT.
D IREC TIO N S FO R US E
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
= 81% ± 3%.
2
en
Non-sterile
5375D-eIFU-0816