User Manual for Pulse Oximeter
Description:
" means the device is on mute status, the
Sound indicator "
user can press recall/mute "
" key to turn on the
indication sound. During the measurement, over-limit event or
probe off event can trigger the alarm.
": Memory full icon; if the memory is full, the memory
"
full icon will appear on the screen. No display of this icon
means the current storing space is not full. If the memory is
full, the data storing will continue in such way the new record
will overwrite the oldest record, so that it's recommended to
upload the stored data into the computer before the memory
gets full.
During the measurement, press recall/mute "
" key to turn
off (or resume) the device sound (including pulse beep sound,
audible alert and key click), while the pulse symbol "
" still
blinks. For alert sound mute function. Refer to Section 6.2 for
detail of the sound mute function.
Note: The pitch tone of pulse beep is modulated by the SpO2
value, that means the pitch tone changes when the measured
SpO2 changes. The higher the SpO2 value is, the higher the
tone frequency of pulse beep (sound becomes sharper); The
lower the SpO2 value is, the lower the tone frequency of pulse
beep (sound becomes flatter).
9