The device is not intended for use in the environment with
rich oxygen or ignitable gas.
Mode of operation: Continuous operation.
Electro-Magnetic Compatibility: Group I, Class B
6 Over-limit Indication
6.1 Limit settings
SpO
low limit setting range: 50% ~ 99%.
2
Pulse Rate limits setting range:
High:
100bpm--240bpm Low: 30bpm--99bpm
During the measurement, if the measured value exceeds the
preset value, the alert beeping sound will be activated, the value
that is over-limit will blink at the same time.
6.2 Over-limit alarm mute setting
During the measurement, if the over-limit alarm is
on, press "
alarm will mute for 90 seconds, but the over-limited
value will still keep blinking. At this moment, the speaker
volume icon becomes " ". If this alert event persists
over 90 seconds, then the over-limit alarm will be
activated again.
During the measurement, if the probe is off or
disconnected, the message "Check Probe" shows on the
User Manual for Pulse Oximeter
" recall/mute key, then the over-limit
23