confirm and exit from date setting.
The procedures of adjusting Month value and Day value
4)
are the same with Year adjustment.
Date Format: yyyy-mm-dd
Note: The setting operations of other parameters (such as TIME,
PATIENT, RECORDING INTERVAL, POWER SAVING etc.)
are the same with date setting.
Time: Time setting
Recording: Time interval for recording data (SpO
there are five options: "1s, 2s, 4s, 8s" and "OFF".
1) "1s": the least length of data record is set to 30 seconds, and
the maximal length for one record is limited to 1 hour. The
total storage time is up to 48 hours.
2) "2s": the least length of data record is set to 60 seconds, and
the maximal length for one record is limited to 2 hours.
The total storage time is up to 96 hours.
3) "4s": the least length of data record is set to 120 seconds,
and the maximal length for one record is limited to 4 hours.
The total storage time is up to 192 hours.
4) "8s": the least length of data record is set to 240 seconds,
and the maximal length for one record is limited to 8 hours.
The total storage time is up to 384 hours.
5) When the option is set to "OFF", the device will not store the
measuring data.
User Manual for Pulse Oximeter
14
& PR),