Technical Specifications; Spo 2 Measurement - CMi PC-66L Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5 Technical Specifications

A. Display Mode: Color dot-matrix LCD;
B. Power Supply:
Supply voltage: d.c. 3.6V (Lithium rechargeable battery)
Operating current: ≤180mA
C. SpO
Measurement
2
Transducer: dual-wavelength LED sensor with wavelength:
Display range: 35~99%
Measuring range: 35~100%
Measuring accuracy:
A
is not greater than 3% for SpO
rms
*NOTE: A
according to ISO 9919 / ISO 80601-2-61.
D. Pulse Rate Measurement
Display and measuring range: 30bpm~240bpm
Accuracy: ±2bpm or ±2% (whichever is greater)
Alarm setting range: 25bpm~250bpm
Default Alarm Setting: High -- 120bpm, Low -- 50bpm
E. Perfusion Index Display
Range: 0.2%~20%
F. Operating Environment
Operating Temperature: 41°F ~104°F
User Manual for Pulse Oximeter
Red light: 663 nm, Infrared light: 890 nm.
Maximal average optical output power:
is defined as root-mean-square value of deviation
rms
range from 70% to 100%
2
21
2mW