18
Technical Information
Installation Procedure
TMP
Agilent
Controller
TMP
Agilent
Controller
Figure 4
1
2
3
4
5
After the master has configured the Profibus network the TMP
Profibus Gateway goes into normal data exchange and the Input and
Output buffer are exchanged between slaves and master.
154/178
Gnd
A
B-
Gnd
A+
B-
Installation
Set the Profibus address by rotational switches on the Profibus
gateway.
Power on the controllers.
Configure the controller to operate in RS-485 mode and set the
address from 1 to 4 (the default is 0), note that the controllers on
the same 485 link must have different addresses.
Power on the ProfiBus interface.
Ready to configure your network, the user can set the
configuration parameters.
TMP Profibus Gateway User Manual / 87-900-976-01 (F)