STERILIZATION
Trays
• Place DePuy Mitek instrument in the appropriate
location within the DePuy Mitek sterilization tray. If a
specific location is not identified for an instrument,
it may be placed in the general purpose area (pin-
mat) ensuring that the sterilant has adequate access
to all surfaces including difficult to reach areas,
lumens, etc.
• Use a validated, properly maintained and calibrated
steam sterilizer.
• Effective steam sterilization can be achieved using
the following cycles:
Cycle Type
Prevacuum
Prevacuum
(outside the U.S.)
Drying Time: 30 minutes
Reassembling the Depth Gauge
To reassemble, align the probe with the body and
insert until fully seated (Figure 2).
Figure 2. Probe and Body Fully Assembled
Temperature
132°C (270°F)
134–137°C
7
Minimum
Exposure Time
4 min
3 min