TECHNICAL SPECIFICATIONS continued
SpO
parameter
2
specifications
Pulse rate parameter
specifications
Pulse strength parameter
specifications:
Resistance to interference of
surrounding light
Resistance to interference of
man-made light
Dimensions (L x W x H)
Net weight with batteries
Classification
JB02007-INS-LAB-RevC16
Measuring range: 35%~99%
Accuracy: ±2% during 75%~99%, ±3% during 50%~75%
Measuring range: 25-250 bpm
Accuracy: ±2 bpm or ±2%
Measuring range: 0.2-20%
Accuracy: ±0.1% during 0.2%~2%, ±1% during 2%~10%, ±2% during 10%~20%
The difference between the value measured in the condition of indoor natural light and
that of darkroom is less than ±1%
Values of SpO
and pulse rate can be accurately measured by pulse oxygen simulator
2
2.59" x 1.42" x 1.3" (66 mm x 36 mm x 33 mm)
2 oz. (60g)
The type of protection against electric shock: Internally powered equipment
The degree of protection against electric shock: Type BF applied part
The degree of protection against harmful ingress of liquids: Ordinary equipment
without protection against ingress of water
Electromagnetic compatibility: Group I, Class B