The following charts show which processing methods each Ipas instrument will withstand.
Summary of Processing Methods for Ipas MVA Instruments
Instrument
Status
when
supplied
by Ipas
Ipas MVA Plus
Clean
aspirator
Ipas Single-
Clean
Valve aspirator
Adapters
Clean
Sterile
Ipas EasyGrip
cannula
(ETO)
Flexible
Sterile
Karman
(ETO)
cannula
Ipas
Sterile
3mm
(ETO)
cannula
† Glutaraldehyde and chlorine are hazardous substances. When processing instruments, take necessary precautions, such
as using personal protective equipment. Refer to the manufacturer's safety instructions to establish safe use.
Minimum
Cleaning
level of
All Ipas instruments
process-
that are reused
ing
should be kept wet
required
until cleaning. A
for use
disinfectant such
as a 0.5% chlorine
solution can be used.
CAUTION: Letting
HLD
the instruments dry
may make it difficult
to completely remove
HLD
all contaminants. To
clean instruments,
wash all surfaces
HLD
thoroughly in warm
water and detergent.
Detergent is preferable
HLD
to soap, which can
leave a residue.
Processing
High-level disinfection
Chlorine †
Boiling
Glutar-
aldehyde †
Yes
Yes
Yes
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Single use only
Single use only

Sterilization
Steam
Glutar-
STERRAD
aldehyde †
autoclave
processor
Yes
Yes
Yes
No
Yes
Yes
Yes
Yes
Yes
Yes
No
No